Section 4: Biomed Functions

Air Detector On/Off

The Air Detector screen can be set to On-High, On-Low, or Off.
WARNING: When the Air Detector is turned off, the pump will not
detect air in the fluid path. Periodically inspect the fluid path and
remove any air to prevent air embolism. Air embolism could result
in serious injury or death to the patient.
1. With the pump stopped and in LL0, access Biomed Func-
tions. (Refer to the beginning of the Biomed Functions
section for instructions on how to access Biomed Functions.)
2. Press
´
3. Use
or
• On-high is the highest sensitivity, where the smallest
bubbles will be detected.
• On-low is lower sensitivity, where only the larger bubbles
will be detected. See Specifications in Section 5.
¤
4. Press
5. Press
46
until "Air Detector" appears.
Î
to select On-High, On-Low, or Off.
to enter the change.
to go to the next screen.