Messages and Alarms
Motor Locked,
remove all power
T
-
WO
TONE ALARM
No Disposable,
clamp tubing
T
-
WO
TONE ALARM
No Disposable,
Pump won't run
T
-
WO
TONE ALARM
Power lost while
pump was on
T
-
WO
TONE ALARM
Programming Incomplete
T
-
WO
TONE ALARM WHEN
STARTING THE PUMP
Reservoir Volume
Empty
T
-
WO
TONE ALARM
Description / Corrective Action
Batteries are depleted and the pump was
powered up with the AC Adapter. Install
new AA batteries, reconnect the AC
adapter, and restart the pump.
The disposable (CADD
Set or Medication Cassette
was removed. Clamp the tubing immedi-
ately. A CADD
®
Administration Set or
Medication Cassette
properly attached in order for the pump
to run. Press
or
alarm.
You have tried to start the pump without
a disposable (CADD
or Medication Cassette
attached. A CADD
or Medication Cassette
be properly attached in order for the
pump to run. Press
silence the alarm.
The pump was on and running when
power was removed. Stop the pump
before changing the battery or removing
the power source. Press
silence the alarm.
A rate or dose must be programmed to
start the pump. Press
silence the alarm.
The Reservoir Volume has reached
0.0 ml. Press
or
alarm. Then install a new fluid container
if appropriate and reset the reservoir
volume.
Section 5: Reference
®
Administration
Reservoir)
Reservoir must be
to silence the
®
Administration Set
Reservoir)
®
Administration Set
Reservoir must
or
to
or
to
or
to
to stop the
51