SIMS Deltec CADD-Legacy PLUS Operator's Manual page 43

Ambulatory infusion pump
Table of Contents

Advertisement

4. Insert a coin into the lock, push in, and turn counterclock-
wise until the line on the lock lines up with the arrow on the
side of the pump and you feel the lock click into place.
WARNING: Attach the cassette (the part of the Medication
Cassette
Reservoir or CADD
attaches to the pump) properly. An improperly attached or
detached cassette could result in unregulated gravity infu-
sion of medication from the fluid container or a reflux of
blood, which could result in death or serious injury to the
patient.
You must use a CADD
Valve or a CADD
gral or an Add On Anti-Siphon Valve to protect against
unregulated gravity infusion that can result from an improp-
erly attached cassette.
5. Gently twist, push, and pull on the cassette to make sure it
is firmly attached. If the cassette is not secure, repeat the
procedure from step 1.
Section 2: Pump Setup and Programming
®
Administration Set that
®
Extension Set with Anti-Siphon
®
Administration Set with either an inte-
1999-01-23 D. Zurn
«Lgc Twist 50 ml»
33