Section 5: Reference
Accuracy Test Results
The following graphs are designed to show flow accuracy of the
infusion system plotted against given time periods.
Flow rate immediately following startup
Time Interval:
Total Time:
Programmed Rate:
Cassette used:
48
36
Flow
24
(ml/hr)
12
0
0
10
Short term flow rate error
Programmed Rate:
Average Flow Rate:
Mean Flow Error:
Cassette used:
% Error
of Flow
62
0.5 min
120 min
24.0000 ml/hr
®
CADD
anti-siphon valve
20
30
40
50
24.0000 ml/hr
23.9227 ml/hr
-0.32%
®
CADD
anti-siphon valve
Administration Set with
60
70
80
90
Administration Set with
100
110
120
120
T (min)
T (min)