Section 5: Reference
Messages and Alarms
[No message]
T
-
WO
TONE ALARM
Error
T
-
WO
TONE ALARM
High Pressure
T
-
WO
TONE ALARM
Key pressed,
Please release
T
-
WO
TONE ALARM
LowBat
T
-
HREE TWO
TONE BEEPS EVERY
50
Description / Corrective Action
With no AC adapter attached, the
batteries have been removed while the
pump is running. The pump is now
stopped and unpowered. Install batteries
to silence the alarm.
Batteries were removed within approxi-
mately 15 seconds after stopping the
pump. Install new batteries to silence the
alarm, if desired. Otherwise, the alarm
will stop within a short period of time.
An error has occurred. Remove the pump
from service and contact Customer
Service to return the pump for service.
The pump has detected high pressure,
which may be resulting from a down-
stream blockage, kink in the fluid path,
or a closed tubing clamp. Remove the
obstruction to resume operation. Or,
press
silence the alarm for 2 minutes, then
remove the obstruction and restart the
pump.
If a key is being pressed, stop pressing it.
If the alarm persists, close the tubing
clamp and remove the pump from use.
Contact Customer Service to return the
pump for service.
The batteries are low, but the pump is
still operable.
5
MINUTES
• Change the batteries soon.
OR
or
to stop the pump and