5.0
Reference

Messages and Alarms, Alphabetical List

Messages and Alarms
Air In Line
Detected
T
-
WO
TONE ALARM
Battery Depleted
T
-
WO
TONE ALARM
Battery Removed
Pump won't run
T
-
WO
TONE ALARM
Description / Corrective Action
The Air Detector has detected air in the
fluid path; the fluid path may contain air
bubbles, or the tubing may not be fully
threaded through the Air Detector. Press
or
to silence the alarm, then:
• Make sure the tubing is threaded
properly.
• If the fluid path contains air bubbles,
close the clamps and disconnect the fluid
path from the patient. Then follow the
instructions for removing air by priming
the pump, described in Section 2. Restart
the pump.
The battery power is too low to operate
the pump. The pump is now stopped.
• Change the batteries immediately.
• Press and hold
pump.
With the AC adapter attached, the AA
batteries have been removed while the
pump is running, or you have tried to
start the pump with depleted batteries.
The pump is now stopped. Press
to silence the alarm. Reinstall
batteries or install new batteries. Press
and hold
to restart the pump.

Section 5: Reference

to restart the
or
49