2.0
Pump Setup and Programming

Installing or Replacing the Batteries

Use new, AA alkaline batteries such as DURACELL
EVEREADY
®
ENERGIZER
grammed values while the batteries are removed.
Dispose of used batteries in an environmentally safe manner, and
according to any regulations which may apply.
WARNING:
• Do not use rechargeable NiCad or nickel metal hydride (NiMH)
batteries. Do not use carbon zinc ("heavy duty") batteries. They
do not provide sufficient power for the pump to operate prop-
erly, which could result in death or serious injury to the patient.
• Always have new batteries available for replacement. If power is
lost, nondelivery of drug will occur and, depending on the type
of drug being administered, could result in death or serious
injury to the patient.
• If the pump is dropped or hit, the battery door or tabs may
break. Do not use the pump if the battery door or tabs are
damaged because the batteries will not be properly secured; this
may result in loss of power, nondelivery of drug, and, depending
on the type of drug being administered, death or serious injury to
the patient.

Section 2: Pump Setup and Programming

®
batteries. The pump retains all pro-
®
or
11