CDM: Programming with Upper Limits, Adjusting
Rate in Lock Level 1
This feature may be used in the continuous delivery mode only. If a
prescription allows for the Continuous Rate to be adjusted during
the course of therapy, you may wish to operate the pump in LL1.
Then, when necessary, you can adjust the Continuous Rate up to
the maximum value that was programmed in LL0.
Programming the pump to use this feature
1. During initial programming in LL0, enter the upper limit
value for the Continuous Rate. (This will be the maximum
value when the pump is in LL1.)
2. After you are finished programming, change the lock level
to LL1.
3. Decrease the Continuous Rate to its starting value, then
¤
press
Adjusting the rate while the pump is in use
If it becomes necessary to increase the Continuous Rate during the
course of therapy, stop the pump but remain in LL1.
1. Press
´
2. Press
3. Restart the pump if appropriate.
Section 2: Pump Setup and Programming
.
until the Continuous Rate screen appears.
Î
or
to select the desired value, then press
¤
.
39