• Do not prime the fluid path with the tubing connected to a
patient as this could result in overdelivery of medication or air
embolism.
• Ensure that the entire fluid path is free of all air bubbles before
connecting to the patient to prevent air embolism.
• Prior to starting infusion, inspect the fluid path for kinks, a
closed clamp, or other upstream obstructions, and remove any
air to prevent air embolism.
Cautions
• Do not operate the pump at temperatures below +2°C (36°F) or
above 40°C (104°F).
• Do not store the pump at temperatures below -20°C (-4°F) or
above 60°C (140°F). Do not store the pump with the Medication
Cassette
™
Reservoir or CADD
the Protective Cassette provided.
• Do not expose the pump to humidity levels below 20% or above
90% relative humidity.
• Do not store the pump for prolonged periods with the batteries
installed.
• Do not immerse the pump in cleaning fluid or water or allow
solution to soak into the pump, accumulate on the keypad, or
enter the battery compartment.
• Do not clean the pump with acetone, other plastic solvents, or
abrasive cleaners.
• Do not expose the pump to therapeutic levels of ionizing radia-
tion.
• Do not expose the pump directly to ultrasound.