Section 4: Biomed Functions
Changing the Delivery Method
Delivery can be changed between Continuous Delivery and Intermit-
tent Delivery. This allows for either continuous drug delivery or
infusion of a prescribed volume of a drug over a specified time period.
When the delivery method is changed, programming will revert to
default settings.
1. With the pump stopped and in LL0, access Biomed Func-
tions. (Refer to the beginning of the Biomed Functions
section for instructions on how to access Biomed Functions.)
„
2. Press
´
3. Press
¤
4. Press
48
until "Delivery Mode" appears.
Î
or
to select "Continuous" or "Intermittent."
to enter the change.