Section 2: Pump Setup and Programming

Watching Power Up

When you install the batteries, the pump will start its power up
sequence during which it performs self-tests and displays pro-
grammed values. Watch for the following:
• Pump model number and last error code ("LEC") if any,
will appear.
• The software version will appear.
• The display will turn on, showing a series of blocks. Look
for any blank areas, which would indicate a faulty display.
• The display will turn off briefly.
• The pump's program screens will appear, followed by
screens showing the Air Detector status, Upstream Occlu-
sion sensor status, and lock level setting. The pump will
beep after each screen. If messages appear, see the Messages
and Alarms Table in Section 5 of this manual for further
explanation and instruction.
• When power up is complete, six beeps will sound, and the
pump will be stopped on the main screen.
NOTE: To move quickly through the power-up screens,
press
Î
press
powered up, it will not respond.
16
repeatedly. To skip the automatic review entirely,
. If you attempt to skip screens before the pump is