Section 2: Pump Setup and Programming
Setting the Lock Level for the Patient
The Lock Level must be changed to LL1 or LL2 to prevent the
patient from having complete access to all programming and oper-
ating functions.
NOTE: You may change the lock level at any time by stopping the
pump and following the procedure below.
To change the lock level
Œ
1. Press
2. The current lock level will appear.
´
3. Press
appears.
Œ
4. Press
´
5. Press
Œ
6. Press
WARNING: Do not disclose to the patient the pump's
security codes or any other information that would allow
the patient complete access to all programming and operat-
ing functions. Improper programming could result in death
or serious injury to the patient.
38
.
Î
or
until the desired lock level (LL1 or LL2)
¤
again or
. "Code 0" will appear.
Î
or
until the Lock Level Code
¤
or
to set the new lock level.
****
appears.