Setting The Lock Level For The Patient - SIMS Deltec CADD-Legacy PLUS Operator's Manual

Ambulatory infusion pump
Table of Contents

Advertisement

Section 2: Pump Setup and Programming

Setting the Lock Level for the Patient

The Lock Level must be changed to LL1 or LL2 to prevent the
patient from having complete access to all programming and oper-
ating functions.
NOTE: You may change the lock level at any time by stopping the
pump and following the procedure below.
To change the lock level
Œ
1. Press
2. The current lock level will appear.
´
3. Press
appears.
Œ
4. Press
´
5. Press
Œ
6. Press
WARNING: Do not disclose to the patient the pump's
security codes or any other information that would allow
the patient complete access to all programming and operat-
ing functions. Improper programming could result in death
or serious injury to the patient.
38
.
Î
or
until the desired lock level (LL1 or LL2)
¤
again or
. "Code 0" will appear.
Î
or
until the Lock Level Code
¤
or
to set the new lock level.
****
appears.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CADD-Legacy PLUS and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

6500

Table of Contents