Section 2: Pump Setup and Programming
IDM: Stopping the Pump During the Dose
Stopping the pump while a dose is in progress will shift all subse-
quent doses by the amount of time the pump is stopped.
Dose
KVO
If you wish to make up for the lost time in the cycle, it is recom-
mended that you wait until the current dose is completed. If the
pump is stopped while a dose is infusing and you reprogram the
"Dose Starts in" screen, then the pump will not deliver the remain-
der of that dose. (To determine the amount of time remaining in the
current dose, use the Dose Remaining screen in the programming
loop.) Then stop the pump, change to LL0 or LL1, and reprogram
the delay on the "Dose Starts in" screen.
Resuming the Dose
To resume delivery of a dose that has been stopped in progress,
simply restart the pump.
Resetting the Cycle
If you wish to reset the cycle, stop the pump during KVO, change to
LL0 or LL1, and reprogram the start of the Next Dose.
If a dose has been stopped in progress and you wish to cancel the
current dose and reset the cycle, make sure the pump is in Lock
40
Pump is
stopped
for 10
minutes
All
subsequent
doses shift by
10 minutes
KVO
Dose
Need help?
Do you have a question about the CADD-Legacy PLUS and is the answer not in the manual?