Section 5: Reference
Messages and Alarms
RUN
ResVol Low
T
HREE SINGLE BEEPS
Service Due
T
-
WO
TONE ALARM
Upstream Occlusion
T
-
WO
TONE ALARM
Value not saved
52
Description / Corrective Action
The Reservoir Volume is low. Change the
fluid container soon. See Reservoir
Volume Alarm in Section 1 for more
information.
Service is due for this pump based on
clock battery age or total motor revolu-
tions. This screen will appear while in
LL0 only for 60 days and then in all lock
levels until returned for service.
Fluid is not flowing from the fluid
container to the pump. Check for a kink
in the tubing or a closed clamp between
the fluid container and pump. Press
or
to stop the pump and silence the
alarm for 2 minutes, then remove the
obstruction and press
pump.
A value was not saved by pressing
Press
to resume programming.
Verify all programming screens before
moving to the next screen or starting the
pump.
to restart the
¤
.