Warnings - SIMS Deltec CADD-Legacy PLUS Operator's Manual

Ambulatory infusion pump
Table of Contents

Advertisement

Read this entire Operator's Manual before operating the
CADD-Legacy
Failure to properly follow warnings, cautions, and instructions
could result in death or serious injury to the patient.

Warnings

• This Operator's Manual should be used by clinicians only. Do
not permit patients to have access to this manual, as the informa-
tion contained would allow the patient complete access to all
programming and operating functions. Improper programming
could result in death or serious injury to the patient.
• For those patients who are likely to be adversely affected by
unintended operations and failures, including interrupted medi-
cation or fluid delivery from the device, close supervision and
provision for immediate corrective action should be provided.
• If the pump is used to deliver life-sustaining medication, an
additional pump must be available.
• The pump is not to be used for delivery of blood or cellular
blood products.
• If the pump is dropped or hit, inspect the pump for damage. Do
not use a pump that is damaged or is not functioning properly.
Contact Customer Service to return a pump for service.
• Use of a syringe with the CADD
in UNDER-DELIVERY of medication. Syringe function can be
adversely affected by variations in plunger dimension and lubric-
ity, which can result in greater force required to move the syringe
plunger. A syringe plunger will lose lubrication as it ages and, as
a result, the amount of under-delivery will increase which could
on occasion, be significant. Therefore, the type of medication
and delivery accuracy required must be considered when using a
syringe with the CADD
Clinicians must regularly compare the volume remaining in the
ambulatory infusion pump.
®
pump.
®
Administration Set may result
iii