3.5
Attaching and Removing the Cassette
3.5.1 Removing a Used Cassette
WARNING:
Alw ays close the fluid path tubing w ith the clamp before removing the cassette
from the pump to prevent unregulated gravity infusion, w hich could result in
death or serious injury to the patient.
To remove the cassette, follow these steps:
Place the pump with the cassette
•
attached in an upright position on a
firm, flat surface.
Insert a coin into the slot in the
•
locking button and turn it one-quar-
ter turn clockwise. The locking but-
ton will pop out when you unlock
the cassette.
Disengage the cassette hooks from
•
the pump hinge pins.
Discard the used Medication
•
Cassette
Reservoir or CADD
™
ministration Set.
NOTE:
For additional inform ation and cautions
associated w ith the M edication Cassette
Reservoir or CADD
refer to the pack age insert w hich
accom panies the product.
22
Ad-
®
™
®
Administration Set,
®
9/20/95 D. Zurn,
E D
LO C K
®