Section 2: Pump Setup and Programming
Programming Screens for Continuous Delivery
These are the programming screens for the
pump, in the continuous delivery mode. Descriptions of the screens
follow.
Reservoir Volume
Continuous Rate
(ml/hr)
Given
Air Detector (Off,
On-High, or On-Low)
Upstream Sensor
(Off or On)
Reservoir Volume
Enter the volume of fluid contained in a filled fluid container. The
Reservoir Volume value decreases as the pump delivers fluid or as
you use prime the tubing. When you change the fluid container,
reset the reservoir volume on this screen. If you do not wish to use
the Reservoir Volume feature, scroll down to "Not In Use" (located
before 1 and after 9999 in the range of values).
The residual volume could be set higher than the capacity of the
fluid container. Be sure to program the reservoir volume to reflect
the actual volume of the medication being used.
20
CADD-Legacy
Reservoir Volume
100.0
Continuous Rate
5.0
Given
2.50
Air Detector
On-High
Upstream Sensor
On
™
PLUS
ml
ml/hr
ml