HBM T10FS Mounting Instructions page 93

Hide thumbs Also See for T10FS:
Table of Contents

Advertisement

Nominal (rated) torque M
Maximum control range
Frequency output
Voltage output
Power supply (version KF1)
Excitation voltage
(square wave)
Calibration signal
triggering
Frequency
Max. current consumption
Preamplifier excitation
voltage
Preamplifier, max. current
consumption
Power supply (version
SF1/SU2)
Nominal (rated) supply
voltage (separated
extra‐low voltage (SELV))
Current consumption in
measuring mode
Current consumption in
startup mode
Nominal (rated) power
consumption
Linearity deviation including
hysteresis, relative to the
nominal (rated) sensitivity
Frequency output
Voltage output
T10FS
N⋅m
100
nom
3
kHz
V
V
V
kHz
A
V
mA
V
DC
A
A
W
%
%
A0785-150 HBM: public
200
500
1 k
4 ... 16
-10.5 ... +10.5 (typ. ±11)
54±5% (peak‐to‐peak)
80±5 %
approx. 14
1 (peak‐to‐peak)
0/0/+15
0/0/+25
18 ... 30; asymmetrical
< 0.9
< 2
< 12
<±0.05 (optional <±0.03)
<±0.07 (optional <±0.05)
Specifications
2 k
3 k
5 k
10 k
93

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T10s2T10s3T10s4T10s5T10s6

Table of Contents