Mechanical installation
Notice
Even if the unit is installed correctly, the zero point
adjustment made at the factory can shift by
approx. ±150 Hz. If this value is exceeded, we advise
you to check the mounting conditions. If the residual zero
offset when the unit is removed is greater than ±50 Hz,
please send the transducer back to the Darmstadt
factory for testing.
For correct operation, comply with the mounting
dimensions (see Page 85).
5.1

Conditions on site

T10FS torque flanges are protected to IP54 according to
EN 60529. They must be protected against coarse dirt
particles, dust, oil, solvents and humidity. During
operation, the prevailing safety regulations for the
security of personnel must be observed (see "Safety
instructions").
There is wide ranging compensation for the effects of
temperature on the output and zero signals of the T10FS
torque flange (see Specifications on Page 91). This
compensation is carried out at static temperatures in
extensive furnace processes. This guarantees that the
circumstances can be reproduced and the properties of
the transducer can be reconstructed at any time.
If there are no static temperature ratios, for example,
because of the temperature differences between the
measuring body and the flange, the values given in the
specifications can be exceeded. So for accurate
measurements, static temperature conditions must then
be obtained by cooling or heating depending on the
T10FS