Soliwave FDR56/FQR56
72

Basic setup

721
Adjustment with free or minimum covered path (function 1)
This automatic setup function can be used if the path is free or minimum covered.
Set encoding switch to Position 1
→ Display of the function number
→ After 2 seconds: Display of the current signal strength, example:
Simultaneously press the operation buttons on the device with free or minimum covered
path
→ Automatic adjustment is carried out
→ Display of the signal strength after adjustment, example:
Move the encoding switch to the initial position 0
→ Display of the current signal strength
Optional:
Additional manual adjustment with maximum covered path (function 4)
For most applications the automatic setup is sufficient. The manual setup can be used to adapt
the microwave barrier individually to the application or medium.
If after an automatic setup on free path (function 1) the medium can not be detected safely
(switching point of the limit detection is not undercut on maximum covered path), the sensiti-
vity must be reduced with this manual setup function 4.
Move the encoding switch to position 4
→ Display of the function number
→ After 2 seconds: Display of the current signal strength, example:
Press the operation buttons on the device to achieve an increase or reduction of the signal
strength display with maximum covered path
→ Display of the signal strength with maximum covered path (all 10 LEDs are not lit)
Move the encoding switch to the initial position 0
→ Display of the current signal strength