Commissioning
5.3

Overview Basic Setup

UB
SD
D
I
LN
L
L
I
E
= empty calibr. (= zero) - setting in 005
F
= full calibr. (= span) - setting in 006
D
= distance interface (distance flange / DK ) - display in 0A6
I
L
= level interface (distance probe end / DK ) - display in 0A5
I
D
= distance complete level - display in 0A5
L
L
= level complete - display in 0A6
L
LN = probe length - setting in 033
UB = upper blocking distance - setting in 059
UP = thickness upper medium
SD = safety distance - setting in 015
20
D
L
DK
UP
L
DK
1
E
F
2
2
1
Levelflex M FMP40 4...20 mA HART
flange:
reference point of
measurement
DL
SD
F
L
E
L
L00-FMP4xxxx-19-00-00-en-038