Commissioning
tank shape
Standard
measuring cond.
Unknown
measuring cond.
Standard
empty calibr.
full calibr.
mapping
safety settings
linearisation
extended calibr.
...
54
6.4

Basic setup

D
LN
L
E = empty calibr. (= zero)
F = full calibr. (= span)
D = distance (distance flange / product)
L
Levelflex M FMP40 FOUNDATION Fieldbus
flange:
reference point of
measurement
UB
SD
setting in 005
setting in 006
display in 0A5
= level
display in 0A6
E
F
LN = probe length
setting in 033
UB = upper blocking distance
setting in 059
SD = safety distance
setting in 015
L00-FMP4xxxx-19-00-00-en-031