General Information

Revision C
Automatic Mode Recording: General Information
In Automatic mode, 12 leads of ECG are acquired simultaneously for a
period of 10 seconds. When initiated with
proceeds automatically.
Depending on the implemented software options, the ECG is:
only printed out (MAC 1200/1200 ST without options MEAS, DIAG),
measured and printed out with the measurement results (MAC 1200/
1200 ST with option MEAS), and
measured, interpreted (analyzed) and printed out with the
interpretative statements (MAC 1200/1200 ST with option DIAG).
127(
The program will only interpret the standard and Cabrera lead
sequences. The NEHB leads will only be measured.
Units equipped with the optional memory function MEMO (MAC 1200/
1200 ST only) can save up to 40 resting ECG. These ECGs can be printed
or sent to CardioSys/CardioSoft (A5 protocol) or to the MUSE CV system
(CSI protocol) (see
"The Storage Program"
The unit offers different report formats for printout of the ECG. With the
system defaults, all 12 leads including the measurement and analysis
results will be documented on a single sheet. (See
page 5-10.)
Some of the system settings can be customized. They are identified with
"configurable."
MAC 1200/1200 ST
2012250-021
, the recording
on page 5-8).
"Report Formats"
on
5-3