Option Code

Revision C
System Setup: Option Code
In this menu you enter option codes to activate a number of optional
software functions. The respective option becomes active after you have
entered the code number. The code numbers are listed on the option code
sheet supplied with the different software options.
1. Use the cursor keys to position the bar cursor on Option Code and
confirm the selection with
The menu for entry of the option code will appear. There is a choice of
seven optional programs.
MEAS: measurement of the 10-second restingECG
DIAG: measurement and interpretation of the 10-second resting
ECG
MEMO: storage of up to 40 resting ECGs
C100:
activates the three options MEAS, DIAG, MEMO for a
maximum of 100 ECGs
C500:
activates the three options MEAS, DIAG, MEMO for a
maximum of 500 ECGs
EVAL
activates the three options MEAS,DIAG, MEMO for a period
of 4 weeks
ERGO stress test (MAC 1200 ST only)
Using the cursor keys, position the bar cursor on the option that you
want to activate.
Enter the 12-digit code number and confirm the entry with
The unit will accept the entered number only if it corresponds to the
unit's serial number. The serial number is indicated at the top of the
menu (Serial Number = xxxxxxxxx). This number must be the same
as printed on the nameplate (back of the device).
When you enter the code number for DIAG and MEMO, the fields for
C100, C500 and EVAL will disappear.
Select
to close the submenu.
MAC 1200/1200 ST
2012250-021
.
.
10-31