Creating Custom Protocols

Revision C
Stress Tests: Creating Custom Protocols
The MAC 1200 ST allows the user to choose among three different test
protocols. The default protocol is the SINGLE protocol with an exercise
phase of 90 seconds. Also available in the list of protocols are the
DOUBLE (180 seconds) and the TRIPLE (270 seconds) protocol.
The protocol can be selected as follows.
1. Press
and select Stress test.
2. Press
. Now the configuration screen is displayed.
3. Move the cursor to "Protocol [others]" and press
available Master's step protocols is shown in the window.
SINGLE
DOUBLE
TRIPLE
You can also edit the parameters of the post-exercise phase as follows.
1. Press
and select Stress test.
2. Press
. Now the configuration screen is displayed.
3. Move the cursor to Protocol [SINGLE] and press
editor is now active.
The protocol editor can be used to change the number of stages in the
post-exercise phase and the duration of the individual stages. You can
define between 1 and 50 stages. The minimum stage time is 30 seconds.
MAC 1200/1200 ST
2012250-021
: the list of the
. The protocol
8-17