Direct ECG Transmission
Selecting the Communication Protocol
Revision C
System Setup: Direct ECG Transmission
Select the Automatic mode and press
Press
to display the configuration menu for the Automatic
mode.
Use the cursor keys to position the bar cursor on Manual copy to
HOST and confirm the selection with
Press
to clear the configuration menu.
Use the cursor keys to position the bar cursor on Communication and
confirm the selection with
Select the same baud rate as at the receiving modem (2400, 4800,
9600, 19200, 38400, 57600).
This adjustment is only necessary with MAC 1200 / 1200 ST units,
because only these devices can send data to CardioSys/CardioSoft or to
the MUSE system. For a MAC 1100 unit that can send data only to
CardioSys/CardioSoft, the communication protocol A5 is preselected.
1. Using the cursor keys, position the bar cursor on Protocol.
2. Select the protocol A5 if you will send data to CardioSys/CardioSoft
system or select CSI to send data to MUSE system.
3. Use the cursor keys to position the bar cursor on Modem, none and
confirm the selection with
MAC 1200/1200 ST
2012250-021
.
([HOST]).
.
.