Location; Date/Time; Lead Fail Beep; High Hr Beep - GE MAC 1200 Operator's Manual

Hide thumbs Also See for MAC 1200:
Table of Contents

Advertisement

Location

*

Date/Time

Lead Fail Beep

High HR Beep

Lead Labels

Pace Enhancement

Baseline Roll Filter

10-22
System Setup: System Setup
3-place ID number of the location to which the ECG recorder is assigned,
default: 1, range: 1 to 600.
Enter date and time (enter 4 digits for the year).
Audio signal to indicate electrode problems (yes, [no]).
Audio signal to indicate that the heart rate exceeds the limit value: yes,
[no] (only in Manual and Arrhythmia modes). The limit value (220 - age)
can be changed manually.
1. The value selected here is the default value appearing in the patient
data
2. with CSI protocol only
[IEC codes]: R, L, F, N, C1 to C6 or
AAMI codes: RA, LA, RL, LL, V1 to V6
Pacer pulses are displayed enhanced and with a given amplitude ([yes],
no).
When programming pacemakers, we recommend to disable the pace
enhancement.
Select the lower cut-off frequency: = 0.04, [0.08], 0.16 Hz.
MAC 1200/1200 ST
2012250-021
Revision C