Continuous Rhythm; Muscle Filter/Ac Filter; Muscle Filter Frequency - GE MAC 1200 Operator's Manual

Hide thumbs Also See for MAC 1200:
Table of Contents

Advertisement

Continuous Rhythm

Muscle Filter/AC Filter

Muscle Filter Frequency

Revision C
System Setup: Automatic Mode
[yes]: In 6_F1 reports, the first 6 leads represent the first half of the 10-
second signal acquisition period, while the second set of 6 leads
represent the 2nd half of the acquisition period.
no:
The second set of 6 leads also represents the first half of the
acquisition period.
The same is true for the report format 3_F1.
127(
When you select Continuous rhythm = yes, the letter R will be
appended to the report name. Report formats 6_F1, 3_F1,
4x2.5R3, 6_F2, H2 become 6_F1_R, 3_F1_R, 4x2.5R3_R, 6_F2_R,
H2_R.
Elimination of muscle artifact and AC line interference.
Muscle filter: [no],
AC line filter: [yes]
127(
Filters may suppress diagnostically relevant portions of the signal,
because they limit the transmission range. Therefore, only enable
filters if necessary.
Cut-off frequency of the muscle filter: [40 Hz], 20 Hz.
The frequency range is indicated in the lower margin of the recording
strip.
0.08 – 40 Hz (40-Hz muscle filter enabled)
0.08 – 20 Hz (20-Hz muscle filter enabled)
0.08 – 150 Hz (muscle filter off).
MAC 1200/1200 ST
2012250-021
10-5