Transmitting Data To A Muse Cv System Via Modem - GE MAC 1200 Operator's Manual

Hide thumbs Also See for MAC 1200:
Table of Contents

Advertisement

Transmitting Data to a MUSE CV System Via Modem

ECG Record Serial Number
Direct Transmission
5-16
Automatic Mode Recording: ECG Transmission
127(
Pacemaker information, telephone number and comments entered in
the patient data are not transmitted to the MUSE CV system.
Before sending data to the MUSE CV system, the MAC 1200/1200 ST
automatically logs on to the database. Then the data transfer is initiated.
If the transfer is stopped, the MAC 1200/1200 ST takes a few seconds
before cancelling the connection because it has to log off the database
first. Then the communication link with the receiving modem is
interrupted and the standard display reappears.
Every ECG record sent to the MUSE CV system contains a unique serial
number for tracking purposes. It is made up of the MAC 1200/1200 ST
serial number + date + time.
127(
The ECG record serial number is printed on the transmission log
only (if enabled). It does not appear on the ECG report printed at the
MAC 1200/1200 ST unit. See
more information.
The unit must be connected to the PC or to the MUSE CV system by
means of the connection cable (pn 223 362 03).
For transmission of the ECG, the unit must be set up as described in
Chapter 10,
"Direct ECG Transmission"
1. After acquisition of the ECG, the transmission is started with
The MAC 1200/1200 ST is also capable of transmitting stored ECGs
(if the memory option MEMO is installed).
Activate the storage program by simultaneously pressing
. (Press the
MAC 1200/1200 ST
2012250-021
"Transmission Log"
section.
button first and hold it depressed.)
on page 10-25 for
.
and
Revision C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for: