10-14
System Setup: Stress Test Mode
Push
to open the Protocol Editor window.
Configuration: Stress Test
Protocol:
[add phase]
add stage
del phase
del stage
print proto
reset proto
Phase
Time In
STAGE
99:00
PRETEST
Ramping
00:00
EXERCISE
Ramping
STAGE 1
2:00
STAGE 2
2:00
STAGE 3
2:00
STAGE 4
2:00
STAGE 5
2:00
STAGE 6
2:00
STAGE 7
2:00
STAGE 8
2:00
STAGE 9
2:00
STAGE 10
2:00
STAGE 11
2:00
STAGE 12
2:00
STAGE 13
2:00
1. A stage time of 99.00 means an indefinite stage time and the program
does not automatically advance to the next stage
2. Values for Ramping are 1 or Yes and 0 or No. Ramping = Yes means the
load increases evenly
3. If the time for the first stage report is longer than the stage time, no stage
report will be printed.
The protocol editor has the form of a table. It lists all parameters of the
protocol: names of the protocol, phase and stage, load parameters
(depending on exercise test device), intervals for the output of reports,
intervals for blood-pressure measurements).
The command menu is shown at the top of the display.
When the editor screen is opened, the cursor highlights the protocol
name (e.g. WHO). With the cursor keys, the cursor can be positioned on
the different commands.
MAC 1200/1200 ST
2012250-021
WHO
add proto
del proto
ok
escape
Basic Lead
1
0
0
0
2
25
50
75
100
125
3
150
175
200
225
250
275
300
325
Protocol Editor Window
Median Repo
First
:
1:50
1:50
1:50
1:50
1:50
3
1:50
1:50
1:50
1:50
1:50
1:50
1:50
1:50