Preparing the product for use

4 pcs. oval head self-tapping screw
1 pc. electric wrist rotator dummy
1 pc. 757L24 charger
1 pc. 10S1=40 lamination ring
1 pc. hook
1 pc. Lamination protection cover (hollow ball)
1 pc. Lamination protection cover (disc with arrow)
1 pc. O-ring
1 pc. clamp ring
4 pcs. mounting brackets
1 pc. foam cover
1 pc. Instructions for use (qualified personnel)
1 pc. Instructions for use (user)
1 pc. Technical information (qualified personnel)
1 pc. 647G563 set of stickers
1 pc. Service pass
6 Preparing the product for use

6.1 Control electronics

The 13E219 control electronics serve to control the elbow joint with up to six MyoBock electrodes and up to two
MyoBock switches or a MyoBock 4-stage control element and one switch. Its flat, slightly vaulted structural shape
permits installation between the inner socket and outer socket.
Control elements with an analogue signal can also be used instead of the electrodes.
6.1.1 Connecting electrodes and/or switches
All control devices have to be connected before installing the control electronics.
INFORMATION
Apply a generous amount of 633F11 silicone grease to all contacts of the control electronics and cable before
connecting them.
Connect the cable connector according to the numbering of the control electronics.
42
MyoBock electrodes
A maximum of six MyoBock electrodes or Ottobock
control elements with analogue signal can be connec­
ted (inputs 1–6).
4-stage control element and switches
A maximum of two switches or one 4-stage control ele­
ment and one switch can be connected.
Connector A: a 4-stage control element or MyoBock
switch can be connected.
Connector B: a MyoBock switch can be connected.
DynamicArm Plus 12K110N=*