Otto Bock DynamicArm Plus Instructions For Use Manual page 58

Hide thumbs Also See for DynamicArm Plus:
Table of Contents

Advertisement

Available languages

Available languages

Handling
Strong signal that begins quickly.
After exceeding the ON threshold, the upper switching
threshold must be reached within 80 ms. The level of
the ON threshold and the upper switching threshold
can be changed manually.
Prog
Activation
3
With two electrodes
Prog
Activation
4
Acceleration with two
electrodes
Prog
Activation
5
Acceleration with a
switch
58
Extension
Control
The speed is proportional to
the control signal.
Control
The longer the control signal
is held, the more the product
accelerates.
Control
Switch position 0: no control
Switch position 1: extension
Switch position 2: flexion
Signal that begins slowly.
The ON threshold is exceeded, but remains below the
upper switching threshold for 80 ms. The level of the
ON threshold and the upper switching threshold can be
changed manually.
Version
Channel
5
or
6
(green
or
brown
marking)
Version
Channel
5
or
6
(green
or
brown
marking)
Version
Channel
5
or
6
(green
or
brown
marking)
Flexion
Switching mode
Flexion:
triggering
channel 5 (green marking)
Extension: triggering with
channel 6 (brown mark­
ing)
Switching mode
Flexion:
electrode
channel 5 (green marking)
The product accelerates
slowly up to the maximum
speed, weak myosignal
above
ON
switching
threshold
Extension:
electrode
channel 6 (brown mark­
ing)
The product accelerates
quickly up to the maximum
speed, strong myosignal
above
ON
switching
threshold
Switching mode
Flexion: channel 5 (green
marking)
The product accelerates
slowly up to the maximum
speed, weak switch signal
DynamicArm Plus 12K110N=*
with
1,
2,

Advertisement

Table of Contents
loading

Table of Contents