Leica CM3050 S Service Manual page 89

Cryostat
Hide thumbs Also See for CM3050 S:

Advertisement

Chapter 2
6. Tighten screws (3).
7. Rotate setscrew (21) further to adjust the
clamping of ball (7).
Coefficient of friction:
Measuring point: outermost point of clamping
lever (23).
8. Tighten setscrew (4).
9. Insert nut (18) into universal joint (17).
10. Insert spindles (14) and (12) with knurled
knobs (13) into the drilling in ball socket (11)
and secure with parallel pin (16).
Make sure both spindles (12) and (14) are po-
sitioned correctly!
Also, both parallel pins (16) must have some
clearance within their guide bush!
If the parallel pins (16) get jammed,
spindles (12) and (14) can no longer be
disassembled, neither can the
clamping of ball socket, front (5) be
adjusted any more!
11. With screws (1) fasten specimen orientation,
assy. to cover (19).
Service Manual / Leica CM3050 S
2.5 kg
Version 1.0 12/00
6. Schrauben (3) festziehen.
7. Durch weiteres Verdrehen des Gewindestiftes
(21) wird nun die Klemmung des Kugelstückes
(7) justiert.
Reibwert : 2,5 kg
Meßpunkt: Äußerster Punkt des Hebels (23).
8. Gewindestift (4) festziehen.
9. Mutter (18) in Kreuzgelenk (17) einstecken.
10. Spindeln (14) und (12) mit Knöpfen (13) in Boh-
rung der Kugelschale (11) einführen und mit Stift
(16) sichern.
Auf korrekte Position der beiden Spindeln (12)
und (14) achten!
Desweiteren müssen beide Zylinderstifte (16) in
ihrer Führung etwas Spiel aufweisen!
Bei Verklemmung der Stifte (16) ist keine
Demontage der Spindeln (12) und (14)
sowie auch keine Justierung der Klem-
mung der Kugelschale (5) mehr möglich!
11. Objektorientierung, kpl. mit den beiden Schrau-
ben (1) am Deckel (19) anbringen.
© Leica Microsystems
2.8 Orientation
Page 51