Download Print this page

Leica CM3050 S Service Manual page 160

Cryostat
Hide thumbs Also See for CM3050 S:

Advertisement

Chapter 3
Access
Press key "KEY", "ARROW UP" and "LIGHT"
(does not work in standby mode)
The following menus can be chosen by pressing the "ARROW" keys.
1. POT = XXX
BRAKE = ZZ
2. WDOG
3. Specimen fast FWD
4. Specimen fast BWD
5. Specimen slow FWD
6. Specimen slow BWD
7. Cutmotor (additional RUN ENABLE-key necessary)
8. set default clock-default adjustment
9. Valve (Cabin) (defrost valve)
10. Magn.Valve Odj. (object cooling)
11. Reg.Valve Obj.
12. Runtime...(in hours) delete with ON
reactive with OFF, resp.by scrolling
13. Heating (Object)
14. Heating (window, pane)
15. Heating (Cabin)
16. Compressor (chamber cooling)
17. Compressor (object cooling)
18. Light
19. Display Light
20. Button LEDs
21. Trim LEDs
22. Clutch
23. Object Position XX
24. Cab.Temp.
25. Temp.
26. ADC Cab. Ref.
27. ADC Obj. Ref
Service Manual / Leica CM3050 S
XXX = tandempot is displayed
ZZ
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
0 – 255 Typical value: 215 + 3
Version 1.0 12/00
© Leica Microsystems
3.17 Service routine
= set brake value
Page 68