Chapter 2
4. Insert cross roller guide rail (I) together with
slider (2).
5. Tighten screws (4) just slightly so they hold
cross roller guide rail (I) in position, and then
center roller cage (V).
6. Setscrews (5) are for adjustment purposes.
The setscrews (5) are tightened using a
dynamometric key (Md = 0,5 Kpcm = 5 Ncm!).
Each time before tightening one of the
setscrews (5) make sure to position the slider (2)
right behind it so that sufficient back pressure is
provided.
7. Tighten screws (4) which hold cross roller guide
rail (I). - Torque = 150 Ncm.
Service Manual / Leica CM3050 S
2.4 Microtome (2.4.5 Cross roller bearings)
4. Kreuzrollenführungsschiene (I) mit Gleitstück (2)
einsetzen.
5. Schrauben (4) zur Befestigung der Kreuzrollen-
führungsschiene (I) leicht anlegen und Rollenkä-
fig (V) vermitteln.
6. Mit den Gewindestiften (5) wird die Einstellung
vorgenommen.
Das Anziehen der Gewindestifte (5) erfolgt mit-
tels Drehmomentschlüssel (Md = 0,5 Kpcm = 5
Ncm!). Dabei muß das Gleitstück (2) immer hin-
ter dem entsprechenden Gewindestift (5) ste-
hen.
7. Schrauben (4) zu der Kreuzrollenführungs-
schiene (I) befestigen. Drehmoment = 150 Ncm.