Leica CM3050 S Service Manual page 67

Cryostat
Hide thumbs Also See for CM3050 S:

Advertisement

Chapter 2
- Check the feed mechanism. For that purpose,
mount dial gauge, part no. 0178 11205 and dial
gauge holder, part no. 0800 35114. Subsequently
(via the electronics), set the feed thickness
values listed below. Perform 10 handwheel
rotations at each setting. After each sequence
of 10 rotations, the following feed values have to
be measured:
1µm
->
14
2µm
->
23
5µm
->
45
10µm
->
90
When doing these measurements, the specimen
cooling must be "OFF".
-
Retraction of the micrometer mechanism
has to be 50µm ± 20µm.
-
Coarse feed travel time 'Fast forward/
backward' from 'Home' to 'Stop' has to be
20 - 24sec.
Service Manual / Leica CM3050 S
- 16µm
- 27µm
- 55µm
- 110µm
Version 1.0 12/00
2.4 Microtome (2.4.6 Micrometer mechanism)
- Überprüfung der Zustellung . Dafür muß die
Mikrometeruhr Part-Nr 0178 11205, sowie der
Mikrometeruhrhalter Part-Nr 0800 35114
montiert werden. Über die Elektronik werden
hiernach folgende Zustellwerte einzeln ein-
gestellt. Bei jeweils 10 Umdrehungen mit
dem Handrad müssen folgende Zustellwerte
gemessen werden.
1µm
->
14
2µm
->
23
5µm
->
45
10µm
->
90
Bei diesen Messungen muß die Objektkopfküh-
lung "AUS" sein.
-
Der Rückzug des Mikrometerwerkes soll
50µm ± 20µm betragen.
-
Die schnelle Vor/Rücklauf-Geschwindigkeit
des Grobtriebes von Home bis Stopposition
soll 20 - 24sek betragen.
© Leica Microsystems
- 16µm
- 27µm
- 55µm
- 110µm
Page 29