Chapter 3
3. Control panel 1 Display (Fig. 23.1, 23.2)
'Preset counter' deactivated
Service Manual / Leica CM3050 S
Upper line:
Immediately after switching on, the instrument
name CM3050 S and the installed software version
are displayed for a short time.
Lower line:
„MOT" for „motorized sectioning on" is not
displayed any longer. The 8 digits on the right of the
lower line (see marked area) are now used to
display a number of additional parameters.
All other settings shown in Fig. 23.1 and 23.2 remain
the same.
By pressing
displayed one after another in the lower line
(sequence as shown on the left). By pressing
the same sequence of parameters is shown,
however in reverse order.
1. Section thickness
Display of selected section thickness in µm. Select
section thickness with
panel 2 while sectioning mode is activated
(Indication TRIM --> OFF).
2. Section counter
Counts the sections carried out.
Press
3. Section thickness totalizer
Displays the total thickness of all sections carried
out. Indication in µm.
Press
4. Preset Counter
To carry out a preselected number of sections.
When sectioning is started, the preset counter
starts counting backwards. When dn=000 is
reached, sectioning in continuous mode stops.
To set back to the preselected value, press button
Press
The preset counter can be deactivated. To activate,
display 'preset counter' pressing the menu button
(see point 4) then press
Prior to starting the preset counter, a value (= pre-
selection of number of sections) must be entered.
Version 1.0 12/00
© Leica Microsystems
3.18 Control elements (supplement)
a number of parameters are
to set back to „0".
to set back to „0".
and
simultaneously.
to set back to „0".
/
buttons of control
to activate.