Leica CM3600 XP
Cryomacrotome
Instruction manual
Leica CM3600 XP
English V 6.1 07/2007
Always keep this manual near the instrument.
Read carefully prior to operating the instrument
Order No.: 0708 35101

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CM3600 XP and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Leica CM3600 XP

  • Page 1 Leica CM3600 XP Cryomacrotome Instruction manual Leica CM3600 XP English V 6.1 07/2007 Always keep this manual near the instrument. Read carefully prior to operating the instrument Order No.: 0708 35101...
  • Page 2 Leica reserves the right to change technical ing thorough investigation in this field. We are specifications as well as manufacturing pro- under no obligation to update the present manual cesses without prior notice.
  • Page 3 ......Dag Graupner December 13, 1995 Managing Director Leica CM3600 XP – Cryomacrotome...
  • Page 4: Table Of Contents

    1.5 Warranty ....................................10 1.6 Service information ................................10 1.7 Decommissioning and disposal ............................10 2. Characteristics of the Leica CM3600 XP ..............12 2.1 Layout of the CM3600 XP ..............................12 2.2 Technical Data ................................... 14 2.2.1 Microtome type ..............................14 2.2.2 Control unit...
  • Page 5: Index

    6.10 Section documentation ..............................77 6.10.1 Tree diagram ................................. 78 6.10.2 Studies protocol ..............................79 6.10.3 Event table ................................. 82 6.11 Photo mode ..................................83 6.12 Screen printout ................................... 83 6.13 Remote inquiry ..................................84 6.14 Help ....................................... 86 Leica CM3600 XP – Cryomacrotome...
  • Page 6 6.17 Foot switch ..................................86 6.18 User log-in / log-out ................................87 6.19 Quitting the application ..............................87 7. Operating the Leica CM3600 XP ..................88 7.1 Check list ..................................... 88 7.2 Switching the instrument on ............................89 7.3 Starting the software ................................ 89 7.4 Setting the chamber temperature ..........................
  • Page 7 Index Leica CM3600 XP – Cryomacrotome...
  • Page 8: Important Notes

    Figures in brackets refer to item num- bers in drawings. Qualification of personnel Only trained laboratory personnel may operate the Leica CM3600 XP. Prior to operating the instrument, the operator must thoroughly read and understand this instruc- tion manual and must familiarize him/herself with all technical details of the instrument.
  • Page 9: Designated Use

    Important notes Designated use The Leica CM3600 XP is an SPS-controlled and motorized precision cryomacrotome for large-surface sections, equipped with a convection- cooled cryochamber for deep-freezing, sectioning, storing and freeze- drying large biological and industrial specimens. The instrument is designed for the above described applications only and must be operated according to the instructions in the present instruction manual.
  • Page 10: Warranty

    Leica Microsystems Nussloch GmbH guarantees that the contractual prod- uct delivered has been subjected to a comprehensive quality control proce- dure based on the Leica in-house testing standards, and that the product is faultless and complies with all technical specifications and/or agreed char- acteristics warranted.
  • Page 11 Important notes Leica CM3600 XP – Cryomacrotome...
  • Page 12: Characteristics Of The Leica Cm3600 Xp

    Characteristics of the Leica CM3600 XP Layout of the CM3600 XP Microtome (in chamber) Emergency stop Window Chamber lid switch Drain tap Knee lever Chamber Port and switch panel Refrigeration Port and switch panel Socket panel and control unit Instruction manual V 6.1 - 07/2007...
  • Page 13 Characteristics of the Leica CM3600 XP Drain tap Condensation water is drained through the drain tap. Refrigeration and control unit The refrigeration and control unit cools the cryochamber and controls the microtome and the entire electronics. Knee lever The knee lever is used for manual operation of the specimen sledge.
  • Page 14: Technical Data

    Knife (160 mm) with facet angle of: 20°, 35° 2.2.2 Control unit The Leica CM3600 XP is equipped with a commercial PC, monitor, key- board, mouse and an internal modem. For further information, please refer to the manual of the PC.
  • Page 15: Dimensions And Weight

    Characteristics of the Leica CM3600 XP 2.2.3 Dimensions and weight Inner (cryochamber L x H x W): 1650 x 720 x 600 mm Outer (L x H x W): 2800 x 1090 x 920 mm (Overall dimensions with connecting cables)
  • Page 16: Refrigeration System - Cooling Chamber

    Characteristics of the Leica CM3600 XP 200 V / 50 Hz, two-phase Nominal voltage: 200 V AC ± 10 % Nominal frequency: 50 Hz Power draw: 2300 VA Automatic cutout/ 2 x 8A / 1 x 10A Mains safety switch...
  • Page 17: General Parameters

    Protective class: Pollution degree: Overvoltage installation category: Instrument parts The Leica CM3600 XP consists of a sliding microtome in a large-volume cryochamber. The instrument is designed for cryosectioning large speci- mens or for processing specimens with the so-called cryo-planing technique.
  • Page 18 The Leica CM3600 XP is operated via PC and/or manually via the knee lever. All data entries are made via the PC. The Leica CM3600 XP features two automatic defrost cycles.
  • Page 19 Characteristics of the Leica CM3600 XP Leica CM3600 XP – Cryomacrotome...
  • Page 20: Safety

    The protective devices on both instrument and accessories must neither be removed nor modified. Only qualified service personnel authorized by Leica may repair the instrument and access its internal components. Warnings The safety devices installed in this instrument by the manufacturer only constitute the basis for accident prevention.
  • Page 21: Safety Standards

    Safety Safety standards The Leica CM3600 XP was built and tested in accordance with the safety regulations for electrical measuring, control, regulating and laboratory devices listed below: • EN 61000-3 -2 : 2000, • EN 61000-3 -3 : 1995 + A1 : 2001, •...
  • Page 22: Knee Lever

    Safety 3.4.2 Knee lever Never leave the instrument unattended when the knee lever is folded out (1) to prevent the microtome from being set in motion acciden- tally. Always swing the knee lever upwards before leaving the instru- ment. 3.4.3 Window The window is heatable and equipped with a lock button (2).
  • Page 23: Special Safety Instructions

    Safety Special safety instructions The safety devices installed in the Leica CM3600 XP by the manufacturer only constitute the basis for accident prevention. Primarily responsible for accident-free operation is above all the institu- tion which owns the instrument and, in addition, the designated personnel who operates, services or repairs the instrument.
  • Page 24 Safety As you work through the instruction manual, familiarizing yourself with the instrument, do not yet install a knife or specimen. Below you find a summary of all warnings contained in this instruction manual. Any use of the instrument not described in this instruction manual is considered improper.
  • Page 25 For safety reasons, we strongly advise you against lifting the microtome (190 kgs) from the cryochamber without appropri- ate lifting gear. Do not attempt to lift the microtome from the chamber manually, even if several persons are there to help. Leica CM3600 XP – Cryomacrotome...
  • Page 26: Site Requirements

    (lowest specified temperature may not be reached, frost may accumulate). All transportations paths for the Leica CM3600 XP must at least be 1.5 meters wide, especially doorways must have that width. 1.5 m is wide...
  • Page 27: Electrical Connections

    (server). Compressed air supply If you want to use the optional extraction system: Compressed air supply for the extraction system (p = 5 - 8 bar) must be available. Leica CM3600 XP – Cryomacrotome...
  • Page 28: Installation

    Installation Unpacking and installation Only duly trained personnel may unpack and install the Leica CM3600 XP. Please contact your local Leica sales organization for appropriate advice. Standard delivery The standard delivery includes: • 1 Leica CM3600 XP, consisting of: • Cabinet •...
  • Page 29: Port And Switch Panel

    External Interface alarm RS 232 signal switch Connection for foot switch Connection for external alarm Connection for Control compressed air lamps Ground fault circuit interruption Mains switch Socket panel Automatic cutout Control lamp Leica CM3600 XP – Cryomacrotome...
  • Page 30: Port And Switch Functions

    Installation 5.3.1 Port and switch functions Mains switch Switch in position “0“ - instrument disconnected from mains except for the socket panel. Socket panel switch Control light is on (red), when the socket panel switch is ON. Foot switch connection Optionally, a foot switch can be connected for the purpose of section thickness documentation (GLP).
  • Page 31: Fuses

    Protects the entire instrument. Automatic cutout Protects the refrigerating sets and all electronic components. 5.3.2 Fuses The Leica CM3600 XP is equipped with the following fuses: • GFCI switch • Automatic cutout • Fuse protecting the mains switch • Fuse protecting the socket panel GFCI switch After the GFCI switch has been activated, it will be in the lower position.
  • Page 32 Installation Fuse protecting the mains switch If the mains switch fuse has blown, please call Technical Service. Fuse protecting the mains switch Fuse protecting socket panel If the socket panel fuse has blown, please proceed as follows: Prior to exchanging the fuse in the socket panel, switch the instrument mains switch off and unplug the instrument from mains.
  • Page 33 Installation The Leica CM3600 XP is equipped with a commercial PC, monitor, key- board, mouse and an internal modem. For further information, please refer to the manual of the PC. The control PC as well as all peripherals connected to it (printer, monitor, etc.) are set to an input voltage of 230V/50Hz.
  • Page 34: Software

    PC must be on and the Microsoft Windows XP ® operating system must have booted and be ready for service. Doubleclicking on Icon „CM3600 XP“ starts the application. Registration mask 1. Type the user name in the identification field. 2. Type the correct password in the password field.
  • Page 35: Description Of The Window Elements

    6.2.2 Tool bar The „Tool bar“ provides quick access (via mouse-click) to individual software functions. When clicking on an icon, the corresponding window will open up. Active icons will light up in green. Leica CM3600 XP – Cryomacrotome...
  • Page 36 Software Icons Clicking on the icon opens the „Main window“, which also opens up after completion of the start and log-in procedure. Clicking on the icon opens the „Parameter settings“ window. Clicking on the icon opens the „Temperature curve“ window. Clicking on the icon opens the „Chronological event list“...
  • Page 37: Work Space

    6.2.4 Information bar The information bar displays the following information: • the currently registered user, • instrument name and software version number, • number of sections carried out so far, • name of laboratory. Leica CM3600 XP – Cryomacrotome...
  • Page 38: Initialization

    Software Initialization To be able to work with the instrument, after completion of the start and log-in procedure the microtome must be initialized. For initialization, proceed as follows: 1. In the „Main window“, click on the INIT button blinking in yellow color (in ‘Mode’...
  • Page 39 INIT button will first change from yellow to green and after a few sec- onds, the caption on the button will change from INIT to AUTOMATIC. The microtome is now ready for service. For further information, refer to chapter 6.4.9. Leica CM3600 XP – Cryomacrotome...
  • Page 40: Main Window

    Software Main window The main window either opens up automatically after comple- tion of the start and log-in procedure or by clicking on the „Main window“ icon. The work space of the main window is divided as follows: • Chamber temperature •...
  • Page 41: Chamber Temperature

    See chapter 6.5.1 ‘Parameter settings’ on how to set the defrost cycle. Manual defrost button Clicking on this button immediately starts a defrost cycle. During the defrost cycle, the button blinks in yellow color. Clicking on the button once more aborts the defrost cycle. Leica CM3600 XP – Cryomacrotome...
  • Page 42: Automatic Dehydration

    Software 6.4.4 Automatic dehydration Indication of start time, start date and duration of the automatic dehydra- tion. See chapter 6.5.1 Parameter settings on how to set the automatic dehy- dration parameters. MANUAL DEHYDRATION button Clicking on this button immediately starts a dehydration cycle. Clicking on this button once more, stops dehydration.
  • Page 43 Clicking on the KNIFE UP button on the right activates the corresponding function. The knife holder will move upwards until reaching the upper limit position. To stop the knife movement, click on the button once more. When activated, the button will light up in green. Leica CM3600 XP – Cryomacrotome...
  • Page 44 Software KNIFE DOWN button on the right Clicking on this button moves the knife holder fast downwards for a short while. If you click on and hold this button, the knife holder will continue to move downward until the left mouse button is released. While active, the button blinks in yellow color.
  • Page 45: Extraction System

    The buttons can be activated only in AUTOMATIC Mode. Clicking on the START/STOP button located in the ‘Mode’ field starts or stops the sectioning program. For further information, refer to chapter 6.4.9. PROGRAM button Clicking on this button activates the set value display. Leica CM3600 XP – Cryomacrotome...
  • Page 46: Sledge Speed

    Software Click on the activated set value field to modify the setting. PAUSE button Clicking on this button stops the sectioning program. Clicking on the button once more restarts the sectioning program, resuming the current section where if was left off. 6.4.8 Sledge speed Indication of current sectioning or return stroke speed in mm/s (millime- ters per second).
  • Page 47: Mode Of Operation

    START/STOP button This button can only be activated in automatic mode, where it starts or stops the sectioning program. When activated the START/STOP and AUTOMATIC buttons will light up in green. Leica CM3600 XP – Cryomacrotome...
  • Page 48: Cutting Window

    Software 6.4.10 Cutting window Indication of the left and right block edge, graphically displayed by blue lines. They indicate the maximum horizontal stroke of the microtome sledge on a scale from 0 - 500 mm. Indication of the actual position of the microtome sledge. The sledge position is graphically displayed by the green line.
  • Page 49: Parameter Settings

    Click on „Parameter settings“ to open the corresponding window. The work space of the parameter settings window is divided as follows: • Parameter setting • Password list • Configuration • Reference voltages • Language selection • Initialization Leica CM3600 XP – Cryomacrotome...
  • Page 50: Parameter Setting

    Software 6.5.1 Parameter setting Set chamber temperature Click on this indication to change the currently selected value. Tempera- ture values from -30 to +1° C can be selected. Alarm temperature Click on this indication to change the currently selected value. Tempera- ture values from -25 to -5°...
  • Page 51 „defrost time“.2 Click on individual parameters (hours or minutes) to change the corre- sponding settings. Click on OK to confirm. Button active/inactive Clicking on this button activates or deactivates the 2nd automatic defrost cycle. Leica CM3600 XP – Cryomacrotome...
  • Page 52 Software Dehydration date Dehydration start time Dehydration duration Indication of start date and time, and duration of the automatic dehydra- tion cycle. Clicking on one of the parameters opens the dialog box „Dehy- dration time“. Click on individual parameters (day, month, year, hour or minutes) to change the corresponding settings.
  • Page 53 Clicking on this button opens the „User configuration’ window. For further information, refer to chapter 6.5.2. EDIT button in the configuration field Clicking on this button opens the „Configuration’ window. For further information, refer to chapter 6.5.3. Leica CM3600 XP – Cryomacrotome...
  • Page 54: Password List

    Software 6.5.2 Password list To access the ‘User configuration’, the registered user must be an administrator. The „user configuration“ indication consists of the following columns: • No. Indication of the program-controlled running number. • Active Indication of status of the user selected. Active users are displayed with an „X“.
  • Page 55 • User name Enter the desired user name. A maximum of 10 characters (numerical or alphanumerical) can be entered. • Full name Enter full first and last name of user. Leica CM3600 XP – Cryomacrotome...
  • Page 56 Software • Access rights Click on the USER RIGHTS line to open a drop down menu. User property ‘Administrator’ or ‘User’ can be selected. Administrators have full access to all application functions. Users do not have access to the password list and to the configuration parameters in the “Parameter settings“...
  • Page 57: Configuration

    • Labels • Comments OK button Clicking on this button saves the entries and closes the „Configuration“ dialog box. CANCEL button Clicking on this button closes the „Configuration“ window without saving any of the changes made. Leica CM3600 XP – Cryomacrotome...
  • Page 58 Software Tabs „Text main study“ and „Text substudy“ The structure of the two tabs is identical. A header and 7 freely selectable comments for each main and/or substudy can be defined. The text entered into the 8 entry fields is copied into the dialog box for creating main and substudies.
  • Page 59 The „General“ tab contains the following entry fields: • Company/lab • Printer for label printing • Font size • Printer for protocoling • Printer for hardcopies • Path for export • Millitron • Activate millitron Leica CM3600 XP – Cryomacrotome...
  • Page 60 Software Company/lab Enter the name of the company or the laboratory. The name will be displayed in the „information bar“, on the right side, see chapter 6.2.4. Printer for label printing Clicking on this line opens a drop down menu. If a printer for label printing is connected, it can be selected in this menu.
  • Page 61 Predefined variables commencing with the special character „@“, represent specific information on the labels. See following page for details. Für die Ausgabe von vordefinierten Variablen, die mit dem Sonderzeichen Leica CM3600 XP – Cryomacrotome „@“ beginnen, können bestimmte Informationen im Etikettendruck ausgegeben...
  • Page 62 Software The following information can be represented as variables: Main study - name Main study - comment line 1 Main study - comment line 2 Current date (date) Main study - comment line 3 Main study - comment line 4 Main study - comment line 5 Main study - comment line 6 Main study - comment line 7...
  • Page 63 Clicking on this indication opens the dialog box ‘Comment on section’. The existing entry can be changed. DELETE button Clicking on this indication opens a safety query dialog. If the safety query is confirmed by clicking on OK, the comment will be permanently deleted. Leica CM3600 XP – Cryomacrotome...
  • Page 64: Reference Voltages

    Software 6.5.4 Reference voltages Indication of the current reference voltage in Volt. Buttons 1 to 8 Clicking on these buttons changes the reference voltage. The individual reference voltage values are predefined and cannot be changed. The display will read the new selected value. Activated buttons will light up in green.
  • Page 65: Temperature Curves

    Click on the temperature curve symbol to open the correspond- ing window. Work space Diagram The window contains the following window panes: • Work space • Diagram The „Temperature curves“ window is only for failure analysis by Leica Service engineers. Leica CM3600 XP – Cryomacrotome...
  • Page 66: Work Space

    Software 6.6.1 Work space REFRESH button Clicking on this button updates the diagram display. RULER button If you click on this button, a ruler will be displayed in the diagram. Use the mouse cursor to move ruler to a certain measuring value position in order to read the measured value more exactly.
  • Page 67 The following check boxes can be activated / deactivated via a mouse click: • T1 to, • T2 to 2, • T3 tv1, • T4 tv Head II, • T5 tv2,I, • T6 tv2,II, • T7 tHg2, • T8 Box. Leica CM3600 XP – Cryomacrotome...
  • Page 68: Diagram

    Software 6.6.2 Diagram The diagram indicates the following: • Temperature in degree centigrade on the y-axis • Time and date on the x-axis • Graphical representation of the 8 temperature measuring points in the instrument by colored lines The diagram can be enlarged as follows: 1.
  • Page 69: Chronological Event List

    The events list is a table displaying the events chronologically, in order of date and time. In this list all events concerning the program are saved. The work space is divided up as follows: • Work space • Event table Leica CM3600 XP – Cryomacrotome...
  • Page 70: Work Space

    Software 6.7.1 Work space FILTER button Clicking on this indication opens the dialog box „Filter“ Here you can define a time filter limiting the event table indication to one day, one week or one month. To set a filter, proceed as follows: 1.
  • Page 71 Number of events displayed in the events list. Print button Clicking on this button creates a screen printout of the current display. Export file button Click on this button to export the events table into a text file. Leica CM3600 XP – Cryomacrotome...
  • Page 72: Event Table

    Software 6.7.2 Event table The table is divided into the following columns: • Date/time Date and time of the events. • User Name of the user who was logged in when the event occurred. • Message Event messages (short description of event). •...
  • Page 73: Alarm List

    Chronological display of the alarm list as a table, in order of date and time. In this list, all events concerning the program are saved. The window contains the following window panes: • Work space • Event table Leica CM3600 XP – Cryomacrotome...
  • Page 74: Work Space

    Software 6.8.1 Work space In the work space the following parameters can be selected: FILTER button Clicking on this indication opens the dialog box „Filter“ For further information, refer to chapter 6.7.1. Comment Clicking on an event in the alarm list activates the ‘Comment’ field. You can enter a comment regarding the selected event.
  • Page 75: Event Table

    For further information, refer to chapter 6.8.1. • User Name of the user who confirmed the alarm. • Message Indication of alarm messages. • Comment Indication of the comment made on the corresponding alarm message. Leica CM3600 XP – Cryomacrotome...
  • Page 76: Error Codes: Meaning And Troubleshooting

    Software 6.8.3 Error codes: meaning and troubleshooting Error code Error message Trouble shooting Error 01 Knife holder - limit position exceed- • Re-initialize instrument. If the problem re- curs, call Technical Service. Error 02 Limit position of sledge exceeded • Re-initialize instrument. If the problem re- curs, call Technical Service.
  • Page 77: Chamber Illumination

    6.10 Section documentation Click on the ‘Section documentation’ symbol to open the corresponding window. Tree diagram Studies protocol Event table The window contains the following window panes: • Tree diagram • Protocol of studies • Event table Leica CM3600 XP – Cryomacrotome...
  • Page 78: Tree Diagram

    Software 6.10.1 Tree diagram Indication of all main studies or substudies in a tree diagram. For naviga- tion purposes, click on the preceding plus or minus sign. The different study types are identified by the preceding colored folder symbols: • Yellow folder symbol Main studies (several).
  • Page 79: Studies Protocol

    2. Select NEW. A dialog box for a new main study or substudy will open 3. Insert the name of the new main or substudy. 4. Press ENTER to confirm your entry and to close the dialog box. Leica CM3600 XP – Cryomacrotome...
  • Page 80 Software When clicking on an existing substudy in the tree diagram the following indication will be displayed in the „studies protocol“ area: The table is divided up as follows: • No. Indication of the running number automatically allocated by the application, in the order of date of creation of the main study.
  • Page 81 Software • Comment Comments can be entered by the user • Knife holder position, Knife holder position at the time the section was made Leica CM3600 XP – Cryomacrotome...
  • Page 82: Event Table

    Software 6.10.3 Event table If you click on a main or substudy in the tree diagram, the following indication will be displayed in the „Event table“: The table is divided into the following columns: • No. Indication of the running event number automatically allocated by the application.
  • Page 83: Photo Mode

    Time until the camera will actually take the photo. For further information, refer to chapter 6.5.1. 6.12 Screen printout Clicking on the ‘Screen printout’ symbol creates a printout of the current screen on a defined printer. For further information, refer to chapter 6.5.3. Leica CM3600 XP – Cryomacrotome...
  • Page 84: Remote Inquiry

    The latest pcAnywhere version will always be displayed. The control PC of the CM3600 XP is equipped with a modem enabling Leica Service Support in case of an instrument malfunction to access directly the control software. The modem connection will be set up by the pcAnywhere software installed on the PC.
  • Page 85 After the connection has been successfully established, the window „Remote Connection Attempted“ opens up. 3. Click on YES to enable access to the PC of the Leica CM3600 XP. Click on NO to deny access. Access to the PC of the Leica CM3600 XP must be confirmed with ‘YES’...
  • Page 86: Help

    Software 6.14 Help Clicking on the „Help“ symbol starts Adobe Acrobat Reader ® and opens the manual in pdf format. 6.15 Saving Click on the „Saving“ symbol to save the section data. ‘Saving’ must be clicked on as a first step to enable section documenta- tion via the foot switch.
  • Page 87: User Log-In / Log-Out

    6.19 Quitting the application Clicking on the „Quit“ symbol closes the application. When quitting the application, all settings made are reset. Any application steps still in progress are terminated. Leica CM3600 XP – Cryomacrotome...
  • Page 88: Operating The Leica Cm3600 Xp

    Operating the Leica CM3600 XP To ensure proper operation of the Leica CM3600 XP as well as use of the specimen stages and the specimen blocks themselves, we strongly recommend to attend a Leica application training. Check list The checklist below provides an overview on how to successfully operate the Leica CM3600 XP.
  • Page 89: Switching The Instrument On

    1. Carry out the software „Start and log-in’ procedure. For further infor- mation, refer to chapter 6.1. If the Leica CM3600 XP had been completely switched off, an initialization procedure must be carried out. For further infor- mation, refer to chapter 6.3.
  • Page 90: Specimen Stages

    Operating the Leica CM3600 XP Specimen stages This chapter provides instructions on how to insert the knives. There are three different types of specimen stages: • Special holder, orientable, with ball-and-socket joint, • Specimen stage, orientable, with ball-and-socket joint, • Specimen stage, non-orientable.
  • Page 91 Operating the Leica CM3600 XP When inserting the specimen stages, bear in mind the following: • Prior to sectioning, the specimen stage should have been stored inside the cryochamber for at least 24 hours, to make sure it has the correct sectioning temperature (ca.
  • Page 92: Inserting The Specimen Stage, Orientable, With Ball-Joint

    Operating the Leica CM3600 XP 7.5.1 Inserting the specimen stage, orientable, with ball-joint To insert the specimen stage, orientable, with ball-and-socket joint proceed as described below. The special holder, orientable, with ball joint is inserted the same way. Tools required: •...
  • Page 93 Operating the Leica CM3600 XP Leica CM3600 XP – Cryomacrotome...
  • Page 94: Inserting The Specimen Stage, Non-Orientable

    Operating the Leica CM3600 XP 7.5.2 Inserting the specimen stage, non-orientable Tools required: • Allen key (size 6). Proceed as follows: 1. Use the knee lever to move the sledge to the right-side reversal point. For further information, refer to chapters 3.4.2 and 6.4.9.
  • Page 95 Operating the Leica CM3600 XP Leica CM3600 XP – Cryomacrotome...
  • Page 96: Knife

    Operating the Leica CM3600 XP Knife This chapter provides instructions on how to insert the knives. Two types of knives can be used in the CM3600 XP: • Standard knives • Disposable blade holder with disposable blade Since there is a considerable risk of injury, please be sure to comply with...
  • Page 97 Operating the Leica CM3600 XP Leica CM3600 XP – Cryomacrotome...
  • Page 98: Inserting Standard Knives

    Operating the Leica CM3600 XP 7.6.1 Inserting standard knives To insert a standard knife, proceed as follows. Tools required: • Allen key (size 4). Proceed as follows: 1. Use the knee lever to move the sledge to the right-side reversal point.
  • Page 99 Operating the Leica CM3600 XP Leica CM3600 XP – Cryomacrotome...
  • Page 100: Inserting The Disposable Blade Holder With Disposable Blade

    Operating the Leica CM3600 XP 7.6.2 Inserting the disposable blade holder with disposable blade To insert the disposable blade holder with disposable blade proceed as described below. Tools required: • Allen key (size 4). • Allen key (size 2.5). Proceed as follows: 1.
  • Page 101 Operating the Leica CM3600 XP Leica CM3600 XP – Cryomacrotome...
  • Page 102: Adjusting The Knife Angle

    Operating the Leica CM3600 XP 7.6.3 Adjusting the knife angle Tools required: • Allen key (size 5). Proceed as follows: 1. Comply with the safety instructions provided in chapter 7.6. 2. Loosen screws (1). 3. Rotate knife holder (2) selecting the desired knife angle on scale (3) (recommended angle = 20°).
  • Page 103: Setting The Trimming Parameters

    Operating the Leica CM3600 XP Setting the trimming parameters Proceed as follows: 1. On the computer screen, open the ‘Main window’. Select the desired section thickness, sledge speed, cutting window and sectioning program. For further information, refer to chapter 6.4.
  • Page 104: Setting The Sectioning Parameters

    Operating the Leica CM3600 XP Setting the sectioning parameters Proceed as follows: 1. On the computer screen, open the „Main window“, select „MANUAL“ mode of operation. Subsequently, select the desired section thickness and the cutting window. For further information, refer to chapter 6.4.
  • Page 105 Operating the Leica CM3600 XP Proceed as follows: 1. Select manual mode of operation and the desired section thickness. 2. Activate the section documentation function. 3. Use the knee lever to move the sledge to the right-side reversal point (left edge of the specimen block just about clears the knife).
  • Page 106: Dehydrating The Sections

    Operating the Leica CM3600 XP 16. Attach the section to a dehydration frame which has already been kept in the cryochamber for a while. 17. To speed up the dehydration process once you are done with the sectioning work, the automatic or manual dehydration function of the instrument can be activated.
  • Page 107 Operating the Leica CM3600 XP Leica CM3600 XP – Cryomacrotome...
  • Page 108: Cleaning And Maintenance

    Cleaning and maintenance Cleaning For cleaning and disinfection of the stainless steel cryochamber and the microtome, use commercial alcoholic disinfectants or alcohol. The outer surfaces of the cryochamber and the control unit can be cleaned with mild household cleaners or with alcohol-based disinfectants or pure alcohol.
  • Page 109 9. Fasten ropes or straps of sufficient lifting capacity to the microtome handles (8). 10. Carefully lift the microtome (5) out of the chamber using the crane. To avoid damages to the cryochamber, make sure that the microtome (5) does not rotate or dangle. Leica CM3600 XP – Cryomacrotome...
  • Page 110: Maintenance

    Cleaning and maintenance Maintenance 8.2.1 Oiling Be sure to check the oil level every and before each section- ing procedure. Checking the oil level Unscrew both lids (1). The oil level must always be above the small capillary tubes (2). Filling up the oil reservoirs Unscrew both lids (1) and add low temperature oil PDP 40 (order no.
  • Page 111: Defrosting

    II must be activated in addition. For further information, refer to chapter 6.5.1. Manual defrost This defrost cycle can be activated manually as needed and starts immediately. For further information, refer to chapter 6.4.3. Leica CM3600 XP – Cryomacrotome...
  • Page 112: Optional Accessories

    (not suitable for use w/other blade holders or instruments) Disposable blades H 45 L, with special coating, 45°, 157 mm length 0216 26153 Sectioning tool Sectioning tool, 150 mm, for Leica CM3600 XP 0417 19039 Instruction manual V 6.1 - 07/2007...
  • Page 113 Freezing box for freezing with liquid nitrogen (-196°C) or with hexane/dry ice (-70°C), capacity of approx. 7 l 0022 50161 Freezing tongs for safe handling of frozen embedding frames and specimen stages 7022 50151 Freezing device for specimen alignment 0022 50152 Leica CM3600 XP – Cryomacrotome...
  • Page 114 Section collecting tape Tesa 4248, can be used for both staining and autoradiography Microtome oils and grease Special oil PDP 40, bottle with 100 ml, for Leica CM3600 XP 0336 19182 Special oil PDP 40, bottle with 500 ml, for Leica CM3600 XP...

Table of Contents