Chapter 2
Disassembly - shaft, assy. (6)
1. Detach screws (14) and remove clutch (15)
together with clutch components (15.1) and
(15.2).
2. Remove dowel pin (13) and then ring (18)
together with washer (7).
3. Remove screws (11) with washers (10).
4. Pull off bearing flange (9). Slide bearing (12) is
friction-set into bearing flange (9).
5. Pull off deep groove ball bearing (8) and washer
(7).
6. Pull out shaft, assy. (6) sideways, then remove
pin (17) and sliding block (5).
7. See chapter 2.4.6 on how to adjust sliding block
(5).
8. To reassemble, proceed in reverse order.
Disassembly - frame (4)
1. Remove screw (3) and spring washer (2).
2. Detach frame (4).
Service Manual / Leica CM3050 S
2.4 Microtome (2.4.7 Shaft with bearing)
Welle, kpl. (6) ausbauen
1. Schrauben (14) öffnen und Kupplung (15) mit
Kupplungsteilen (15.1) und (15.2) abnehmen.
2. Spannstift (13) entfernen und Ring (18) mit
Scheibe (7) abnehmen.
3. Schrauben (11) mit Scheiben (10) entfernen.
4. Lagerflansch (9) abziehen. Gleitlager (12) ist in
Lagerflansch (9) eingepresst.
5. Kugellager (8) und Scheibe (7) abziehen.
6. Welle (6) seitlich herausziehen, Stift (17) und Ku-
lissenstein (5) entfernen.
7. Justierung des Kulissensteins (5), siehe Kapitel
2.4.6.
8. Die Montage erfolgt in umgekehrter Reihenfolge.
Gestell (4) ausbauen
1. Schraube (3) und Federscheibe (2) entfernen.
2. Gestell (4) abnehmen.
Version 1.0 12/00