W&H teon + RID-200 Operating Instructions Manual page 97

Disinfector
Table of Contents

Advertisement

Setting the temper-
ature and time
Setting temperat-
ure 1
Setting drying time
1
Setting temperat-
ure 2
The drying phase consists of two blocks. The temperature and dura-
tion (holding time) must be set for each block.
The first block (temperature 1 and drying time 1) is not assigned in all
programmes, but if necessary can be set up by the W&H Customer
Service Department.
   ...
Temperature 1
         
The temperature is set in 5 ° increments. The possible range is shown
in the bottom line of the display.
Use the arrow buttons  (higher) and  (lower) to set the temperat-
ure.
Press OK to save the setting.
   ...
Drying time 1
         
The holding time is set in 1 minute increments. The possible range is
shown in the bottom line of the display.
Use the arrow buttons  (higher) and  (lower) to set the holding
time.
Press OK to save the setting.
   ...
         
Temperature 2
The temperature is set in 5 ° increments. The possible range is shown
in the bottom line of the display.
Use the arrow buttons  (higher) and  (lower) to set the temperat-
ure.
Press OK to save the setting.
Programme settings
Temperature 1
75
°C
(50 - 115
°C)
Drying time 1
30
Min
Min)
(0 - 120
Temperature 2
75
°C
(50 - 115
°C)
97