W&H teon + RID-200 Operating Instructions Manual page 65

Disinfector
Table of Contents

Advertisement

Resetting the
counter
Setting the interval
The counter for the cleaning interval may be reset only after clean-
ing has been completed.
Reset
Yes
No
Yes
The counter is reset.
No
The counter will not be reset.
Select an option using the  and  arrow buttons and confirm your
choice with OK.
The interval depends on the number of programme sequences and
must be set on the basis of usage and the expected number of
particles/solids in the soiling.
Example of tubular filter:
For weekly cleaning with 2 programme sequences per day and 5
working days in the week, this yields an interval of 10 (2 x 5 = 10).
With a higher incidence of particles, a shorter interval should be se-
lected in order to clean the tubular filter several times weekly. With a
lower incidence of particles, weekly cleaning is sufficient.
We recommend cleaning the tubular filter after every 10 programme
sequences.
The setting value is entered in increments of 5. The possible range is
shown in the bottom line of the display.
Use the arrow buttons  (higher) and  (lower) to set the interval.
Press OK to save the setting.
Interval
10
(5 - 100)
Settings 
65