Settings
HEPA filter re-
placement
Cleaning the fil-
ters in the wash
cabinet
Cleaning the
A 800 tubular filter
Activating and
setting the inter-
val
64
Filter maintenance
The air filter in the drying system must be replaced regularly with a
new one. For more information on changing the filter, see "Mainten-
ance / Changing the HEPA filter".
The filters in the wash cabinet must be checked and cleaned daily,
see "Maintenance/Cleaning the filters in the wash cabinet".
A counter in the controls can be activated to remind you of the re-
quired cleaning at regular intervals.
The A 800 tubular filter can be used in special injector bars on various
mobile units and baskets and must be cleaned at least once a week.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
For the weekly cleaning, a counter in the controls can be activated to
remind you of the required cleaning at regular intervals.
Open the menu as follows:
button
Settings
Filter maintenance
Filter combination
Filter combination
Active
Inactive
Active
–
The cleaning interval is activated.
The
Active
selection allows you to reset the counter or set the
cleaning interval.
–
Inactive
The cleaning interval is deactivated.
Select an option using the and arrow buttons and confirm your
choice with OK.
Tubular filter
or
Tubular filter
Active
Inactive