Plumbing
AD water connec-
tion for
30-1,000 kPa (UK:
100-1,000 kPa)
pressure - pres-
sure-resistant
DI water ring line
IMPORTANT
146
This washer-disinfector is supplied as standard for a pressurised sys-
tem operating between 30-1,000 kPa (UK: 100-1,000 kPa). If the wa-
ter pressure is below 200 kPa the water intake duration will be auto-
matically increased.
The pressure tested hose for AD water, coded green, has a ¾ inch
threaded union for connection to the onsite stopcock for AD water.
 If the washer-disinfector is not going to be connected to demin-
eralised (DI) water, the DI water connection has to be deactivated
by the W&H Customer Service Department. The intake hose re-
mains on the back of the machine.
The machine can be connected to a ring line system for DI water. For
this purpose, it must be technically adapted and the controls reset by
the W&H Customer Service Department.
Please contact the W&H Customer Service Department for further
information.
UK, Australia and New Zealand.
This appliance must be installed according to AS/NZS 3500.1 (Aus-
tralia and New Zealand) or in accordance with WRAS regulations
(UK). This appliance has been supplied with a separate backflow
prevention device.
This machine must be connected to the potable water supply via the
non-return valve (check valve) supplied with the machine.
Before making plumbing connections, ensure the appliance is discon-
nected from the mains power supply (switch off or unplug from the
power supply).
Turn off the mains water tap.
Place the seals on both sides of the non-return valve.
Connect the female end of the non-return valve to the mains water
tap (3/4" thread).
Connect the filter to the male end of the non-return valve (3/4"
thread).
Connect the inlet hose to the filter.
Ensure that all connections are screwed into position correctly. The
connection point is subject to mains water pressure.
Turn on the tap slowly and check for leaks.
Correct the position of the seal and union if necessary.