Allocating wash
blocks

Altering a programme

A programme is altered in two steps:
– Altering the programme begins with a list of all wash blocks alloc-
ated to the programme. First this list must be confirmed.
– Then the programme parameters can be altered.
Use this option to document all changes to factory settings in case
of a subsequent Service call requirement.
Programme and dispensing changes must be documented for val-
idated processes. In Germany this is a requirement of the Medical
Devices Operator ordinance (MPBetreibV).
The process must be validated again as necessary.
   ...
Programme settings
         
Change programme
             
Change programme
Select the programme you want to alter.
For more information see "Allocating wash blocks".
Every programme change begins with a list of the wash blocks.
Vario TD Dental
Accept
All wash blocks which are allocated to the programme are listed in
the display. The allocation can be adapted as necessary by W&H Ser-
vice.
Select the option
Further setting options will then follow. You can edit these in any or-
der you want.
Vario TD Dental
Vario TD Intensive
Pre-wash 1
Accept
and confirm with OK.