Programme
header

Adjusting programme settings

The programme settings should be adjusted to suit technical require-
ments and the load.
Additional specialist knowledge is required to alter programme set-
tings and this should therefore be undertaken only by experienced
users or by the W&H Customer Service Department.
Programme and dispensing changes must be documented for val-
idated processes. In Germany this is a requirement of the Medical
Devices Operator ordinance (MPBetreibV).
The process must be validated again as necessary.

Programme structure

Each programme is subdivided into programme blocks which run one
after another. A programme consists of at least one and a maximum
of eleven programme blocks. Each block can occur only once in a
programme.
The so-called programme header is placed above the programme
blocks and contains general programme settings. Individual wash
block parameters are also globally activated or deactivated here.
Change volume of water
Every programme block with water intake is allocated a nominal
water quantity. The water quantity can be raised or lowered incre-
mentally for all blocks to the base value in the programme chart.
Drain time
If the on-site waste water system is insufficient to drain the waste
water from the wash chamber within the time allocated, the drain-
age time can be lengthened by a set amount.
Parameters for measuring wash pressure and spray arm monitoring
can only be accessed by the W&H Customer Service Department.

Programme settings

89