Adding and dispensing chemical agents
Checking con-
sumption

Refill indicator

Dispensing neut-
ralising agent
50
Check consumption regularly by checking the fill levels in the supply
containers and replace containers in good time to avoid the dispens-
ing system being sucked completely dry.
When the fill level is low in the
agent you are reminded to refill it.
Confirm the message shown with OK and
Refill the neutralising agent as described.
If it has run out, the machine will be locked for further use.
It will be ready for use again when the supply container has been
replaced.
The dispensing concentration is set by the W&H Customer Service
Department.

Instrument care products

 Instrument care products based on paraffin oils (white oils) can
damage elastomers and plastic in the machine.
Such care products must not be dispensed as chemical agents in
this machine even if they are recommended for machine use by the
care product manufacturer.
If required, paraffin oil based instrument care products can be used
after machine reprocessing within the scope of instrument care. Ob-
serve the specifications of the manufacturers of the instruments and
care products.
It is acceptable however to reprocess instruments which have been
treated with such care products in this machine.