Unsatisfactory Cleaning And Corrosion - W&H teon + RID-200 Operating Instructions Manual

Disinfector
Table of Contents

Advertisement

Problem solving guide

Unsatisfactory cleaning and corrosion

Problem
There are white deposits
on the wash load.
The load is flecked.
128
Cause and remedy
The water softener is set too low.
Set the water softener to the correct water hardness.
There is no salt in the salt reservoir.
Refill the reactivation salt.
The quality of the water for the final rinse was insufficient.
Use water with a low conductance value.
If the machine is connected to a water softening cart-
ridge, check it and replace as necessary.
The water from the AD water connection is not sufficiently
softened.
Check the pre-selected water softening units. If neces-
sary, replace the water softening cartridge with a new
one.
The rinsing agent container is empty.
Refill the container.
The rinsing agent concentration is set too low.
Contact the W&H Customer Service Department and
have the dispensing concentration reset.

Advertisement

Table of Contents
loading

Table of Contents