Refill Indicator; Dispensing Rinsing Agent - W&H teon + RID-200 Operating Instructions Manual

Disinfector
Table of Contents

Advertisement

Adding and dispensing chemical agents

Refill indicator

Dispensing
rinsing agent
48
Close the container.
Wipe up any spilled rinsing agent. This prevents over-foaming oc-
curring during the next programme.
When the fill level is low in the (DOS 2) supply container for rinsing
agent you are reminded to refill it.
Confirm the message shown with OK and
refill the rinsing agent as described.
The dispensing concentration is set by the W&H Customer Service
Department.
– If there are spots of water left on load items after reprocessing, then
the dispensing concentration is set too low.
– If clouding or smearing appears on load items, the dispensing con-
centration is set too high.
In either case contact the W&H Customer Service Department and
have the dispensing concentration reset.