Operation
5.6.6
Determines one or several functions that are executed when a trigger event occurs.
The CFO100 series provides you with one trigger input 1 (IN 0), the CFO200 series with four parameteriz-
able trigger inputs 1 ... 4 (IN 0 ... IN 3).
Here, four different functions can be assigned to each trigger input to four different states.
The following trigger events are possible: Level High, Level Low, Rising edge, Falling
Trigger inputs can be used for the following purposes: Switching outputs, teaching or removing colors or
color groups, removing color table, running autogain (automatic modulation) of the receiver or activating key
lock.
Each trigger can be linked to one or several of the above activities, see also more differentiated list of func-
tions, see Fig.
Trigger inputs can be used for the following purposes:
- Switching the outputs
- Teaching colors as new group
- Adding colors to current group
- Teaching and grouping colors continuously
- Deleting color table
- Automatic modulation for sensor LED
- Rerecording white reference
- Activating key lock
Each trigger can be linked to one or several of the above activities, see also more differentiated list of func-
tions, see Fig.
colorSENSOR CFO

Trigger Input

Go to the menu Settings > Trigger input 1 ... 4.
50.
51.
edge, see Fig.
49.