Appendix 1 Technical Specifications; A1.1 Safety Specifications - EDAN SE-1 User Manual

Electrocardiograph
Table of Contents

Advertisement

SE-1 Electrocardiograph User Manual

Appendix 1 Technical Specifications

A1.1 Safety Specifications

Standards:
Anti-electric-shock type:
Anti-electric-shock degree:
Degree of protection against
harmful ingress of water:
Disinfection/sterilization
method:
Degree of safety of application
in the presence of flammable
gas:
Working mode:
EMC:
Patient
Leakage
Current:
Patient
Auxiliary
Current:
Dielectric Strength:
IEC 60601-1:2005/A1:2012
EN 60601-1:2006/A1:2013
IEC 60601-1-2:2007
EN 60601-1-2:2007/AC:2010
IEC 60601-2-25:2011
Class І with internal power supply
Type CF
Ordinary equipment (Sealed equipment without liquid proof)
Refer to the user manual for details
Equipment not suitable for use in the presence of flammable
gas
Continuous operation
CISPR 11 Group 1, Class A
NC
<10μA (AC) / <10μA (DC)
SFC
<50μA (AC) / <50μA (DC)
NC
<10μA (AC) / <10μA (DC)
SFC
<50μA (AC) / <50μA (DC)
4000V rms
- 52 -